The FEI Titan G2 80-200 CREWLEY is a remotely operated field emission gun (scanning) transmission electron microscope equipped with a high-brightness Schottky field emission electron gun, a monochromator unit, a Cs probe corrector, a Super-X EDX system, an UltraScan 1000XP-P charge coupled digital camera, an on-axis bright-field/dark-field STEM detector as well as a Gatan Enfinium ER (model 977) spectrometer with DUAL EELS option allowing a simultaneous read-out of EDX and EELS signals at a speed of 1000 spectra per second.

The Titan G2 80-200 CREWLEY is characterised by a STEM resolution of 80 pm and TEM information limit of 110 pm at 200 kV. Basic specifications for the instrument are given below.

Microscopists planning to use this instrument should contact András Kovács or Roland Schierholz.

Users of the Titan G2 80-200 CREWLEY are kindly asked to quote a tecnical description of the instrument published in the Journal of large-scale research facilities 2 (2016) A43. http://dx.doi.org/10.17815/jlsrf-2-68 when referring to the use of this instrument in publications.

specifications
  • Acceleration Voltage
80 kV ... 200 kV
  • X-FEG Brightness (200 kV)
1.8 E9 A/cm2/sr
  • Information Limit (TEM @ 200 kV)
1.1 Å
  • Point Resolution (TEM @ 200 kV)
2.4 Å
  • Combined Image and Sample Drift (TEM @ 200 kV)
< 0.3 nm/min
  • Astigmatism Instabliity (TEM @ 200 kV)
< 0.6 nm/min
  • Resolution (STEM @ 200 kV and 50 pA)
0.8 Å
  • EDX System Energy Resolution (10 kcps)
< 136 eV @ MnKa
  • EDX Solid Angle
0.9 sr
  • Combined Electron Probe and Sample Drift
< 0.3 nm/min
  • EELS System Energy Resolution (120 and 200 kV)
0.65 eV
detectors
  • Gatan UltraScan 1000XP-P (2k x 2k) Charge Coupled Device Camera.
  • HAADF STEM Detector
  • On-axis Triple DF1/DF2/BF Detectors
  • FEI Super-X EDX System with 120 mm2 Combined Detector Area
  • Gatan Enfinium ER (modell 977) Spectrometer
specimen stages
  • Double Tilt Low Background Holder
± 35°
  • Specimen Drift DTLBH
cf. above
  • High Field of View Single Tilt Tomography Holder
± 70°

   
© by the er-c | legal notice | data protection | er-c@fz-juelich.de